Odor Efficiency Testing: Principles, Sampling and Interpreting the Results

Odor can be measured. An olfactometry test shows how efficiently an odor control installation works, but only if the samples are taken, transported and analyzed according to the standard.

W
WIZZUM engineering teamPublished on 03/02/2026 · Updated on 16/11/2026 · 5 min read

CONTENTS

IN SHORT

  • Odor concentration is measured by dynamic olfactometry according to EN 13725, in a laboratory accredited to ISO/IEC 17025.
  • Samples are taken in Nalophan bags filled to 80–90% and analyzed the same day if possible, and within 30 hours at most.
  • The result in ouE/m³ must always be read together with the measurement uncertainty. Ignoring it is a common and serious mistake.

Why measure odor

Odor is often treated as a subjective complaint from neighbors. In fact, it is an environmental parameter that can be measured. Olfactometry tests are used to evaluate odor control installations, such as scrubbing towers, and help on three fronts: regulatory compliance, installation performance and responsibility toward the surrounding area.

For the plant it has a practical use. The measurement shows whether the scrubber or other odor control installation really reduces the odor, and provides hard data for discussions with the authorities or neighbors, instead of one side’s opinion against the other’s.

Standards and accreditation

  • Odor concentration is measured by dynamic olfactometry according to EN 13725. The current edition is EN 13725:2022, which replaced the 2003 version.
  • The laboratory must meet the general requirements of ISO/IEC 17025:2017.
  • The analyses are carried out by an external accredited laboratory (in Spain, accredited by ENAC).

Sampling

ParameterRequirement
Sampling pointScrubber inlet, before treatment. Where needed, the result is compared with a measurement at the outlet.
BagsNalophan (PET), 10, 25 or 55 l
TubingPTFE, 8 mm diameter
FillingApproximately 80–90% of the volume, without filling them completely
Transport and storageBelow 25 °C and above the dew point, in opaque containers and protected from light
Time to analysisIdeally the same day; 30 hours at most

The bags are not filled completely because the pressure changes during transport, especially if the samples travel by air. The bag and tubing materials meet the requirements of EN 13725.

All these conditions have the same aim: that the laboratory receives the same air that was sampled at the installation. Excessive temperature, light or water vapor condensing inside the bag can alter the sample, and then the result no longer reflects how the scrubber really performs.

Olfactometry tests: from sampling to the result in ouE/m³Olfactometry tests: from sample to resultDynamic olfactometry according to EN 13725 (current version: 2022)1Sampling• Inlet and outletthe installation• Nalophan bagsof 10, 25 or 55 l• Filled to 80–90%• 8 mm PTFE tubing2Transport• Below 25 °C,above the dew point• Protected from light• Analysis, ideallythe same day;max. within 30 h3Laboratory• AccreditationISO/IEC 17025• Dynamicolfactometry• EN 13725 standard4Result• Odor concentrationin ouE/m³• Always togetherwith the measurementuncertaintyHow to assess odor abatement efficiencyInlet (before treatment)odor concentration, ouE/m³Outlet (after treatment)odor concentration, ouE/m³comparisonA difference smaller than the measurement uncertainty does not necessarily indicate a real change at the installation.Source: EN 13725, ISO/IEC 17025; procedure described in the article.wizzum.com
Diagram: olfactometry test from sampling to the result in ouE/m³.

Laboratory analysis

The samples are analyzed by an independent accredited laboratory using dynamic olfactometry. The main result is the odor concentration in European odor units per cubic meter (ouE/m³).

If samples have been taken at the inlet and outlet, comparing the two odor concentrations shows how efficiently the installation removes odor.

The result has to be interpreted together with the measurement uncertainty interval described in EN 13725. Ignoring the uncertainty is a common and serious mistake when assessing an installation's efficiency.

In practice, this means that two results that differ by less than the measurement uncertainty do not necessarily indicate a real change in how the installation is working.

Additional analyses

If the situation calls for it, the olfactometry test is complemented with other measurements:

  • pollutants: Dräger colorimetric tubes in the field or chemical analysis in the laboratory;
  • VOCs: real-time measurement with a MiniRAE 3000 meter or laboratory analysis;
  • TOC: determination of total organic carbon in the laboratory. It helps to understand where the odor comes from and how efficiently the treatment works.

Final report

The report presents the odor concentration results and, if additional measurements have been made, also those for pollutants, VOCs and TOC. A good report ensures sample traceability, compliance with the standards, clear treatment of uncertainty and a technical justification that will stand up before the authorities or an auditor.

You can come back to that report after upgrading the installation, to compare the new results using the same criteria.

Conclusion

The reliability of the result depends on three things: rigorous sampling, analysis in an accredited laboratory and correct interpretation of the uncertainty. When all three are in place, dynamic olfactometry provides objective evidence of how the installation affects odor in the surrounding area. We discuss emission measurements further in the article on air and gas measurements.

Frequently asked questions

How much time can pass between sampling and analysis?+

Ideally, the sample reaches the laboratory the same day. EN 13725 allows a maximum of 30 hours between sampling and analysis.

Why aren't sampling bags filled completely?+

They are filled to 80–90% because the pressure changes during transport, especially by air, and the bag needs some spare volume.

What units is the result given in?+

In European odor units per cubic meter (ouE/m³), always together with the measurement uncertainty.

W

AUTHOR

WIZZUM engineering team

We design and commission wastewater and gas treatment installations for industry, from benchmarking and FEED engineering to start-up.

NEED HELP?

Do you want to check how your odor control installation is performing?

  • Comparative technical analyses
  • Design and delivery of turnkey installations
  • Environmental benchmarking and BAT analysis

Environmental engineering for industry.
Industrial wastewater, emissions and odors.

OFFICE

Avda. Cabrera, 36
08302 Mataró (Barcelona), Spain
info@wizzum.com

© 2026 Wizzum. All rights reserved.